Bump Tests and Calibrations Failing
Why are my bump tests and calibrations failing?
Calibrations and bump tests can fail if there is not enough gas reaching the sensors.
- To ensure that gas is flowing from the gas
cylinder to G6 Dock, check that:
- G6's gas sensor filter is clean.
- There are no kinks, crimps, or leaks in the tubing.
- The gauge on the gas cylinder indicates there is gas inside (i.e., check that the gas cylinder is not empty), and the gas tank valve is open.
- The gas cylinder is using a demand-flow regulator.
- The quick-connects are properly connected to G6 Dock.
- The gas tubing is connected to the correct inlet.
- Gas concentrations in the gas cylinder match dock configurations.
- There is no ambient gas present.
- If G6 Dock is still not operating correctly, please contact Technical Support.