Bump Testing and Calibrating

To manually bump test and calibrate G6, you will need a calibration cap (ACC-G6-CAL) and tube (ACC‑G6‑T2). You can also bump test and calibrate using G6 Dock.

  1. Attach the tubing to the calibration cap. Ensure the other end of the tubing is attached to a fixed flow regulator on the gas tank. Do not turn on the gas cylinder until G6 indicates you should do so.
  2. From the Home screen, select by pressing the center button.
  3. Using the navigation buttons, scroll through the Global Navigation menu and select Compliance ()by pressing the center button ().
  4. To access the available Compliance procedures, select by pressing the right button.
  5. Using the navigation buttons, scroll to and select Bump test or Calibration by pressing the center button ().
  6. Follow the screen prompts to complete the test.

For detailed information on G6, see our G6 Technical User Manual.