Manual Calibration
For a manual calibration, you need:
-
A cylinder or cylinders containing the appropriate gases
-
A 0.5 LPM (or 1 LPM for Cl2, HCN, NO2, NH3, SO2, or H2S) fixed-flow regulator attached to the cylinder(s)
-
A tube fitted with a quick connect coupling insert
If you are calibrating EXO Diffusion in a windy environment, you must use the wind guard. See the Using the Wind Guard section for more information on using the wind guard.
If EXO displays a "Calibration fail" message:
-
Check the gas and cylinder connections.
-
Check that cylinder gas concentrations match EXO calibration gas configuration.
-
Let EXO sit until its gas readings stabilize.
-
Try the calibration again.
If the calibration continues to fail, contact your organization’s safety supervisor.
If you know EXO is in a clean atmosphere and a gas sensor is reading abnormal levels, this can mean the sensor’s baseline has shifted and EXO’s displayed gas readings are not accurate. Try calibrating the sensor. If the sensor is still reading abnormal levels, you may need to zero the sensor.