Gas Cylinders

You can manually calibrate sensors at the same time using one gas cylinder or individually using multiple gas cylinders. If using multiple cylinders, manual calibration must be repeated for each cylinder. In step 6 of the manual calibration procedure ( Manual Calibration section), ensure you select the gas sensors that correspond to the attached cylinder's gas mixture. The gas concentration of the connected gas cylinder should match the gas concentration listed in EXO's calibration gas configuration in Blackline Live.

Note:

Some cartridges require you to calibrate sensors in a specific order due to gas sensor cross sensitivity. For more information, see G7 and EXO Gas Sensor Bump Testing and Calibration Order.