Zeroing G6

How to zero G6 sensors to reset the baseline.

If configured for your device, you can manually zero your sensors to reset the baseline if G6 is not reading zero and you know you are in an atmosphere with no gas.

Important:

If G6’s baseline appears to have shifted, it is best to calibrate your sensor. If you cannot perform a calibration, zero your device.

Note:

The baseline reading for oxygen (O2) is 20.9 %vol.

Warning:

Only zero G6’s gas sensor in a known clean environment.

  1. From the Home screen, select by pressing the center button.
  2. Using the navigation buttons, scroll through the Global Navigation menu, then select Compliance by pressing the center button ().

    The Compliance screen opens, displaying when your device’s next bump test and calibration are due.

  3. To access the available Compliance procedures, select by pressing the right button.

    The available Compliance procedures display.

  4. Using the navigation buttons, select Zero sensor by pressing the center button ().
  5. To start the zero workflow, select by pressing the right button.
    Note:

    To cancel and exit the workflow, press the left button.

    G6 zeros the sensor.

  6. Once the zeroing successfully completes, select by pressing the right button to return to the Compliance screen.
Important:

If the zero fails, acknowledge the event to exit the current workflow and retry the operation.

Note:

If G6 is currently experiencing a sensor error alarm event, you cannot zero your device. Select the center button to exit the workflow.

Contact your organization's safety professional or Blackline Safety Technical Support for assistance troubleshooting your device.