Operational notifications include:

  • Bump test due (optional)

  • Calibration due (optional)

  • Bump test reminder (optional)

  • Calibration reminder (optional)

  • Timer done

  • Low battery

  • Lost connection

  • Pairing needed (G7x only)

Low battery

The low battery notification interval is configurable (10%-70%) by your Blackline Live administrator. Your Blackline Live administrator can also mute the notification light, sound, and vibration patterns associated with this notification.

The low battery notification is activated when G7 detects that it is operating below the configured low battery threshold. G7 remains in low battery status until it is charged above the configured threshold.

Lost connection

The lost connection interval (in minutes) is configurable by your Blackline Live administrator.

The lost connection notification is activated when G7 fails to connect with Blackline Live for the configured time interval (e.g., five minutes).

Bump test due

The bump test due notification is activated when G7 is overdue for a bump test.

Following the bump test due notification, a message will display in the banner until the device is successfully bump tested. The device remains fully functional.

Bump test reminder

The bump test reminder notification interval is configurable by your Blackline Live administrator. Your Blackline Live administrator can also mute bump test due notification lights, sounds, and vibrations.

The Bump test reminder notification is activated when G7 is due for a bump test within the interval (hours or days) configured in Blackline Live.

Calibration due

The calibration due notification is activated when G7 is overdue for calibration.

Following the calibration due notification, a message will display in the banner until the device is successfully calibrated. The device will remain fully functional.

Calibration reminder

The calibration reminder notification interval is configurable by your Blackline Live administrator. Your Blackline Live administrator can also mute calibration due notification light, sound, and vibration patterns.

The calibration reminder notification is activated when G7 is due for calibration within the interval (hours or days) configured in Blackline Live.

Timer done

The timer done notification is activated when the G7 timer counts down to zero (0).

Pairing needed (G7x only)

The pairing needed notification is activated when G7x does not have a network key to connect to G7 Bridge.

For more information on connecting G7x to G7 Bridge, see the G7 Bridge Technical User Manual.