Setting Calibration and Bump Test Reminders
Enabling the calibration and bump test reminders feature ensures that you are notified that a calibration or bump test is due upon start-up instead of while you are in the field.
For example, if you have a daily bump test interval, your bump test is due every 24 hours.
If you have not set up a reminder window and you start your shift two hours early one day, you are not prompted to bump test your device because 24 hours has not elapsed. When the bump test becomes due, you may already be out in the field without bump test materials.
If you have set up a two-hour bump test/calibration reminder window and start your shift two hours early, G7 notifies you that the bump test is due on start-up because the bump test is due within this two-hour window. As a result, you avoid a bump test due event while in the field.
The bump test and calibration reminder window can be customized from G7’s configuration profile in Blackline Live. For more information, contact your Blackline Live administrator.