Use the Gas info option to view G6's gas settings.

Use the Gas info option to view your device’s gas settings, including:

O2 DevicesH2S, SO2, CO Devices
  • Gas sensor (O2)

  • High enrichment setpoint (%vol)

  • Low enrichment setpoint (%vol)

  • Baseline (%vol)

  • Low depletion (%vol)

  • High depletion (%vol)

  • Peak enrichment (%vol)

  • Peak depletion (%vol)

  • Last bump test (date)

  • Last calibration (date)

  • Calibration gas concentration (from your device’s configuration profile) (%vol)

  • Gas sensor (H2S, SO2, or CO)

  • High gas setpoint (ppm)

  • Low gas setpoint (ppm)

  • Peak gas reading (ppm)

  • STEL calculation (ppm/15 min)

  • TWA calculation (ppm/8 hrs)

  • Last bump test (date)

  • Last calibration (date)

  • Calibration gas concentration (from your device’s configuration profile) (ppm)

Note:

G6 setpoints are configured in Blackline Live. G6’s default high and low setpoint ranges are described in the G6 Default Gas Setpoint Ranges table. For more information on configuring high and low gas setpoints, contact your Blackline Live administrator.

Table 1. G6 Default Gas Setpoint Ranges
Gas Default Setpoint Range Increments
H2S0.5 – 50 ppm0.1 ppm
SO2 0.5 – 100 ppm0.1 ppm
CO5 – 500 ppm1 ppm
O2 0.1 – 25 %vol0.1 %vol

G6 gas sensor display ranges are from 0 ppm/%vol to the sensor maximum. Gas readings above sensor maximums result in over limit (OL) values. For detailed information about G6 gas sensor ranges, seeGas Sensor Specifications.