Calibrating G6
Calibrating G6 ensures that the gas sensor can accurately detect gas levels throughout its operating life.
If configured for your device, you can manually calibrate your gas sensor by applying a known concentration of gas for a set amount of time.
This procedure ensures the gas sensor can accurately detect gas levels throughout its operating life. The calibration schedule should be based on your company’s safety policy.
For more information on G6 calibration notifications, see Operational Notifications.
To manually calibrate G6, you need a calibration cap (ACC-G6-CAL) and tube (ACC‑G6-T2). You can also calibrate using G6 Dock.
For more information on calibrating your device with G6 Dock, see the G6 Dock Technical User Manual.
Only perform calibrations in a known clean environment. G6 does not monitor for environmental gas during calibration.