Automatic firmware installs occur exclusively when G6 is removed from G6 dock or when a manual bump test, calibration, or sensor zero are performed.

Automatic firmware installs take place exclusively when a G6 is removed from G6 Dock or when a manual bump test, manual calibration, or manual sensor zero are performed, as the device is assumed to be in gas-free environment where maintenance operations can be performed.

Important:

While docked, G6 has a continuous network connection. After a docked bump test or calibration, G6 automatically verifies if a new version of firmware has been downloaded onto the device. If so, G6 initiates the firmware install process after the device is removed from G6 Dock.