Installing the EXO Pump components
Ensure that the EXO Diffusion components have been uninstalled. For detailed instructions, refer to Uninstalling EXO Diffusion Components.
- Install the pump gas chamber:
- Using your fingers, attach the pump gas chamber coupling to EXO’s inlet tubing.
- Carefully push the inlet tubing back into EXO’s interior.
- Slide the pump gas chamber into its slot and secure the chamber using two screws and washers, ensuring they are snug but not over-tight.
- Install the pump module:
- Insert the gas sensor cartridge that was previously removed by sliding it into its slot in the pump gas chamber.
- Place the pump cartridge cover on EXO, ensuring that you properly orient the cover over top of the front enclosure. Tighten the cover’s screws, ensuring they are snug but not over-tight.
EXO Pump is ready for operation.
|
Note:
EXO Pump’s pump inlets must be assigned before using the device. For detailed instructions on configuring pump inlets, refer to the EXO Technical User Manual. Important:
Although EXO Pump inlet filters are an optional accessory, Blackline recommends always installing inlet filters to maintain the pump in good condition. | |