Wind Guard

When bump testing or calibrating EXO Diffusion in windy conditions, the Wind Guard is required to help regulate the flow and concentration of gas delivered to the sensors.

  1. Position the Wind Guard in front of EXO with the narrow ends on top, as shown below.
  2. Slide the Wind Guard into the front air diffusion vents (underneath the Top Lights). When the Wind Guard is in the proper position, the two clamps will lock.
  3. Bump test or calibrate as usual.
  4. When bump test or calibration is complete, immediately remove the Wind Guard and store it in a clean, dry environment for later use.
Warning:

The Wind Guard MUST be removed when the bump test or calibration is complete. Failure to remove the Wind Guard will slow down EXO’s response to hazardous gases.