Wind Guard
When bump testing or calibrating EXO Diffusion in windy conditions, the Wind Guard is required to help regulate the flow and concentration of gas delivered to the sensors.
- Position the Wind Guard in front of EXO with the narrow ends on top, as shown below.
- Slide the Wind Guard into the front air diffusion vents (underneath the Top Lights). When the Wind Guard is in the proper position, the two clamps will lock.
- Bump test or calibrate as usual.
- When bump test or calibration is complete, immediately remove the Wind Guard and store it in a clean, dry environment for later use.
Warning:
The Wind Guard MUST be removed when the bump test or calibration is complete. Failure to remove the Wind Guard will slow down EXO’s response to hazardous gases.
