Help Center

Search the Help Center

Gas Sampling

EXO Pump has four inlets. When connected to tubing, the inlets allow one EXO to monitor multiple remote areas. Inlets must be assigned before they are functional. For more information, see the EXO Technical User Manual.

  1. Power on EXO and press OK to open the main menu.
  2. Use the arrows and OK button to select Gas options.
  3. Select Inlet settings. By default, the pump inlets (1-4) will display as OFF.
  4. Using a tube fitted with a quick connect coupling insert, attach the tube to the desired inlet.
  5. Select Inlet settings, and then select the inlet you connected the tube to in step 4.
  6. Use the arrows and OK button to toggle the inlet to ON. This pump inlet is now functional.
Important:

If you start the pump while operating at temperatures of -20°C (-4°F) or lower, EXO will generate a pump blocked alarm that persists until the inlet warms up and begins operating normally. In a multiple inlet sampling cycle, the inlets do not have time to warm up. For temperatures below -20°C (-4°F), use only single gas sampling.