Manual Bump Test
To perform a manual bump test, you need:
-
A cylinder or cylinders containing the appropriate gas(es)
-
A 0.5 LPM (or 1 LPM for Cl2, HCN, NO2, NH3, SO2, or H2S) fixed-flow regulator attached to the cylinder(s)
-
A tube fitted with a quick connect coupling insert
If you are bump testing EXO 8 Diffusion in a windy environment, you must use the wind guard. See the Using the Wind Guard section for more information on using the wind guard.
If a bump test fails:
-
Check the gas and cylinder connections.
-
Let EXO sit until its gas readings stabilize.
-
Try the bump test again.
-
If the bump test continues to fail, contact your organization’s safety supervisor.