Gas Cylinders
You can manually bump test gas sensors at the same time using one gas cylinder with a multi-gas mixture or individually using multiple gas cylinders. If using multiple cylinders, you must repeat the manual bump test procedure for each cylinder. In step 6 of the manual bump test procedure (Manual Bump Test section), ensure you select each gas sensor that corresponds to the attached cylinder’s gas mixture. The gas concentration of the connected gas cylinder should match the gas concentration listed in EXO ’s calibration gas configuration in Blackline Live.
Some cartridges require you to bump test sensors in a specific order due to gas sensor cross sensitivity. For more information, see G7 and EXO Gas Sensor Bump Testing and Calibration Order.