Bump Test Schedule
You can configure the bump test schedule to match your company’s safety policy. These changes are made in the configuration profile in Blackline Live.
Blackline recommends that your bump test interval not exceed 30 days. Where site or regulatory requirements are more stringent, Blackline recommends the more stringent requirements apply.
The bump test results in either a pass or fail. Exiting the bump test before all sensors have been tested results as a fail.
EXO displays when the next bump test is due during start-up and on the Home screen. By default, an overdue bump test causes a reminder message to display in the banner of the Gas status screen. The bump test overdue reminder is configurable in Blackline Live.