Uninstalling EXO 8 Diffusion Cover and Chamber

When converting EXO from the diffusion configuration, you will also need to remove the diffusion cartridge cover, the cartridge, and diffusion chamber.

  1. Power off EXO.
  2. Use a Phillips screwdriver to remove the diffusion cartridge cover, then set the cover aside.
    Note: The captive screws should stay attached to the diffusion cartridge cover for storage.
  3. Remove the device's gas sensor cartridge by pulling it out of its slot in the diffusion chamber, then set it aside.
  4. Remove the screws and washers from each side of the diffusion chamber.
  5. Slide the chamber out from its slot, then use your fingers to carefully detach the diffusion chamber coupling from EXO's inlet tubing. Set the diffusion chamber aside.
    The inline fitting may stay attached to the diffusion chamber coupling. If this happens, remove the fitting from the diffusion chamber coupling and reconnect it to the internal tubing.
    Note: You may need to gently unclip or adjust the tubing by sliding it through the tube clamp to shorten the loop and allow for the movement of the diffusion gas chamber.
    CAUTION: During any adjustment of the inlet tubing and clamp, take care not to dislodge the display ribbon cable connector. If you dislodge the connector, gently reconnect it before proceeding.