Calibration Gas Requirements
Calibration Gas Selection
The sensor is specifically designed to measure methane concentration and should be calibrated using a methane-based calibration gas.
Calibration Gas Composition
For accurate and reliable calibration, follow these guidelines:
The calibration gas should ideally contain 50% LEL methane.
The gas mixture must not contain other combustible gases, as these may interfere with calibration accuracy.
Multi-component mixtures (e.g., Quad and Quint gas) may be used, provided methane is the target combustible gas.
The balance gas can be either air or nitrogen.