Hydrogen Cyanide (HCN) and Nitrogen Dioxide (NO2) Sensor Cross-Sensitivity Technical Notes
Cross sensitivity is a phenomenon that occurs when a gas other than the gas being monitored causes the gas sensor to show a reading, even when the target gas is not present. An example of this would be a CO sensor responding to H₂ in the area.
Most cross sensitive responses cause a sensor to read a higher concentration than the actual concentration of its target gas. The interaction between the hydrogen cyanide (HCN) and nitrogen dioxide (NO2) electrochemical sensors works differently. Exposure to one gas can cause the other sensor to read an under limit reading, which is a reading below the normal zero baseline.
If a device or multiple devices within one area include both an HCN sensor and an NO2 sensor, exposure to one of these gases can cause the other sensor to temporarily stop displaying a valid reading for its own target gas.
For more information about cross-sensitivity, see Gas Sensor Cross Sensitivity.
HCN and NO2 Cross-Sensitivity Testing
Blackline Safety's Engineering team tested the HCN and NO₂ sensors cross-sensitivity by applying each gas individually at 10 ppm for 5 minutes and observing the response of the other sensor.
| Gas Applied | Sensor Affected | Observed Response | Recovery After Exposure Ends |
|---|---|---|---|
| HCN | NO2 sensor | Sensor drops below its normal zero baseline (under limit) | The sensor took approximately 292 seconds to begin recovering, then briefly overshot to about 0.5 ppm before settling back to 0 ppm after roughly another 103 seconds (about 6.5 minutes total). The sensor then responds normally to NO2. |
| NO2 | HCN sensor | Sensor drops below its normal zero baseline (under limit) | Sensor impacted for approximately 5 seconds. The sensor then responds normally to HCN. |
Once the interfering gas cleared and the affected sensor recovered, both sensors responded normally to its own target gas. The cross-sensitivity response window is during and immediately after exposure, when the cross sensitive sensor is reading under limit rather than reflecting real readings for its target gas.
Why This Matters for Your Sensor Configuration?
This HCN and NO2 cross-sensitivity is a property of each sensor's own electrochemical cell, not of the cartridge or the device it is installed on. That means the same effect can occur whether the HCN and NO₂ sensors are in the same multi-gas cartridge on one device or installed on two separate devices that are both monitoring the same atmosphere. In either case, if both HCN and NO₂ are present in the surrounding air, each sensor can be driven under limit by the other gas.
If your application involves an atmosphere where both HCN and NO₂ may be present, contact Blackline Safety's Technical Support team or your account representative before finalizing your sensor or cartridge configuration to discuss what is best for your organization.
Bump Testing and Calibration
During bump testing or calibrations, expect that applying HCN calibration gas may cause a temporary under limit reading on the NO₂ sensor, and applying NO₂ calibration gas may cause a temporary under limit reading on the HCN sensor. This is expected behavior and not a device fault. Wait for the recovery times noted in the table above before relying on subsequent readings from the impacted sensor.
Sensor Behavior During Events
If you observe a low or high gas notification on one sensor followed by a sudden drop to a zero or under limit reading on the other sensor, do not assume the hazard has cleared. Follow your work site's evacuation and escalation procedures based on the triggered alert, and report the event to Blackline Safety's Technical Support team so we can help investigate.
Technical Support
Contact details for the Blackline Safety Technical Support team.
Need help? Our Technical Support team is available 24/7 to assist you.
North America
Toll Free: 1-877-869-7212 | mailto:support@blacklinesafety.com
United Kingdom
+44 1787222684 | mailto:eusupport@blacklinesafety.com
International
+1-403-451-0327 | mailto:support@blacklinesafety.com