Gas Sensor Specifications

GAS CARTRIDGE AND SENSORS
GASSensor TypeRangeResolution
Ammonia (NH3)Electrochemical0-100 ppm0.1 ppm
Carbon Dioxide (CO2)NDIR0-50,000 ppm50 ppm
Carbon Monoxide (CO)Electrochemical0-500 ppm1 ppm
Chlorine (Cl2)1Electrochemical0-20 ppm0.1 ppm
Chlorine Dioxide (ClO2)1Electrochemical0-2 ppm0.01 ppm
COSH (CO & H2S)Electrochemical0-500 ppm CO and 0-200 ppm H2S1 ppm CO and 0.1 ppm H2S
High-Range Ammonia (NH3)Electrochemical0-500 ppm1 ppm
High-Range Carbon Monoxide (CO)Electrochemical0-2000 ppm1 ppm
HIGH-RANGE HYDROGEN SULFIDE (H2S)Electrochemical0-500 ppm0.5 ppm
Hydrogen (H2)Electrochemical0-40,000 ppm400 ppm
Hydrogen Cyanide (HCN)Electrochemical0-30 ppm0.1 ppm
Hydrogen Fluoride (HF) 1Electrochemical0-10 ppm0.1 ppm
Hydrogen Resistant Carbon Monoxide (CO-H)Electrochemical0-500 ppm1 ppm
Hydrogen Sulfide (H2S)Electrochemical0-100 ppm0.1 ppm
LEL - Infrared (LEL-IR)4NDIR0-100% LEL1% LEL
LEL- Molecular Property Spectrometer (LEL-MPS)3MPS0-100% LEL1% LEL
Nitrogen Dioxide (NO2)Electrochemical0-50 ppm0.1 ppm
Oxygen (O2)Pumped electrochemical0-25% vol0.1% vol
Ozone (O3) 1Electrochemical0-1 ppm0.1 ppm
Photoionization (PID)PID0-4,000 ppmVariable2, lowest 0.1 ppm
Sulfur Dioxide (SO2)Electrochemical0-100 ppm0.1 ppm

1This sensor is not compatible with the pump cartridge.

2Dependent on correction factor.

3Operating pressure: 80 to 120 kPa (11.6 psi to 17.4 psi).

4 The A-CART only operates in pump mode, regardless of configuration.

Note:

To avoid gas sensor cross-sensitivity responses, confirm sensor combinations with Blackline Safety Technical Support .