Configuring Operating Modes

Operating modes are customized in the G7 configuration profile in Blackline Live. Each profile supports up to a total of five modes. These modes allow G7 to temporarily change its behavior for different situations and can be turned on and off through G7’s interface.

Available operating modes include:

  • Normal mode

  • Pre-entry

  • SCBA

  • Leak check

  • Pump run

  • High risk

  • Over LEL

Note:

Some modes can be configured in Blackline Live so that they are only available when a pump cartridge is installed. To enter a mode that uses the pump, you must complete a successful pump block test.

Note:

If the Auto pump start configuration is enabled or A-CART is installed, G7 devices will only operate in pumped modes. Diffusion modes are disabled, including:

  • Normal operation (pump off)

  • Pre-entry (pump off)

  • High risk

  • Leak check (pump off)

  • SCBA

  • Over LEL

Important:

Operating modes used in areas with potential gas (pre-entry, SCBA, leak check, and over LEL) have a timeout period. After the timeout interval elapses, G7 prompts you whether you would like to continue in this mode:

  • If you select Yes, the mode remains active.

  • If you select No, G7 returns to normal operation.

  • If you do not make a selection within 30 seconds, G7 automatically returns to normal operation. If you have a check-in timer enabled, G7 prompts you to check in.

For more information on gas detection configuration modes, see the Blackline Live Technical User Manual.