Intrinsically Safe Certification
Intrinsically Safe This device is certified Intrinsically Safe gas detector for use in Class I Division 1 Groups A,B,C,D T4; Ex ia IIC T4 Ga; Class I Zone 0 AEx ia IIC T4 Ga hazardous (classified) locations. Improper use of device may impair protection provided by the device. CSA23CA80191458 IECEx CSA 23.0047 CSANe 23ATEX1211 MC267256
Cl I Div 1 Grp A,B,C,D T4 Cl I Zn 0 AEx ia IIC T4 Ga Ex ia IIC T4 Ga-20°C ≤ Ta ≤ +55°C Warning: Charge only in a nonhazardous location. This equipment may only be charged by a power supply unit with a limited energy electric circuit in accordance with CAN/CSA C22.2 No. 61010-1-12 and ANSI/UL 61010-1, or Class 2 as defined in the Canadian Electrical Code C22.1, Section 16-200 and/or National Electrical Code (NFPA 70), article 725.121. The maximum voltage and current from the charger shall not exceed 5.625Vdc and 2A respectively. | Ce dispositif est un détecteur de gaz certifié intrinsèquement sûr pour une utilisation dans les endroits (classés) dangereux de Classe I Division 1 Groupes A,B,C,D T4 ; Ex ia IIC T4 Ga ; Classe I Zone 0 AEx ia Groupe IIC T4 Ga. Une utilisation inappropriée du dispositif est susceptible d'altérer la protection qu'il fournit. Standards: CAN/CSA C22.2 No. 60079-0: 2019 CAN/CSA C22.2 No. 60079-11: 2014 UL 913, Eighth Edition UL 60079-0: Seventh Edition UL 60079-11: Sixth EditionEN 60079-0: 2018 EN 60079-11: 2012 IEC 60079-0: 2019 7th Edition IEC 60079-11: 2011 6th EditionWarning: Chargez uniquement dans un en- droit non dangereux. Cet
équipement ne doit être chargé qu'avec une alimentation utilisant un circuit
électrique à énergie limitée conformément aux normes CAN/CSA C22.2 No.
61010-1-12 et ANSI/UL 61010-1, ou à la Classe 2 telle que définie dans le Code
canadien de l'électricité C22.1, Section 16-200 et/ou le Code national de
l'électricité (NFPA 70), article 725.121. La tension et l'intensité maximales du chargeur
ne doivent pas dépasser 5,625 V CC et 2 A respectivement. |