Gas Sampling

EXO 8 Pump configuration has four inlets. When connected to tubing, the inlets allow a single EXO to monitor multiple remote areas. Inlets must be assigned before they are functional.
  1. Power on EXO and press the center button to open the Main menu.
  2. Use the left, right, and center button to select Gas options.
  3. Select Inlet settings. By default, the pump inlets (1-4) display as OFF.
  4. Using a tube fitted with a quick connect coupling insert, attach the tube to the desired inlet.
  5. Select Inlet settings, and then select the inlet you connected the tube to in step 4.
  6. Use the left, right, and center button to toggle the inlet to ON. This pump inlet is now functional.
Important:

If you start the pump while operating at temperatures of -20°C (-4°F) or lower, EXO generates a pump-blocked notification that persists until the inlet warms up and begins operating normally. In a multiple inlet sampling cycle, the inlets do not have time to warm up. For temperatures below -20°C (-4°F), use only single-gas sampling.