Gas Sampling
- Power on EXO and press the center button to open the Main menu.
- Use the left, right, and center button to select Gas options.
- Select Inlet settings. By default, the pump inlets (1-4) display as OFF.
- Using a tube fitted with a quick connect coupling insert, attach the tube to the desired inlet.
- Select Inlet settings, and then select the inlet you connected the tube to in step 4.
- Use the left, right, and center button to toggle the inlet to ON. This pump inlet is now functional.
Important:
If you start the pump while operating at temperatures of -20°C (-4°F) or lower, EXO generates a pump-blocked notification that persists until the inlet warms up and begins operating normally. In a multiple inlet sampling cycle, the inlets do not have time to warm up. For temperatures below -20°C (-4°F), use only single-gas sampling.
