Installing G7 EXO Diffusion Components
To upgrade G7 EXO Diffusion, or to convert from G7 EXO Pump to G7 EXO Diffusion, you will install the diffusion components.
You will need:
-
G7 EXO with the previous components removed
-
G7 EXO Diffusion components or G7 EXO Diffusion Upgrade kit (ACC‑G7EXO‑DIFFUSION-MOD-KIT)
-
Screwdriver (#2 Phillips)
The previous G7 EXO Diffusion or G7 EXO Pump components must be completely uninstalled before proceeding.
- Install the diffusion chamber:
- Using your fingers, attach the diffusion chamber coupling to G7 EXO’s inlet tubing.
- Carefully push the tubing into G7 EXO’s front enclosure.
- Slide the diffusion chamber into its slot and secure the chamber using two screws and washers, ensuring they are snug but not over-tight.
- Install the diffusion module:
- Insert the device’s gas sensor cartridge by sliding it into it’s slot in the diffusion chamber.
- Place the diffusion cartridge cover on G7 EXO, ensuring that you properly orient the cartridge cover over top of the device’s front enclosure. Tighten the cartridge’s screws, ensuring they are snug but not over-tight.
G7 EXO Diffusion is ready for use.
Note:
When bump testing or calibrating G7 EXO Diffusion in windy conditions, the wind guard is required to help regulate the flow and concentration of gas delivered to the sensors. Warning:
Always remove the wind guard as soon as G7 EXO has been successfully bump tested and calibrated. Failure to do so will slow down EXO’s response to hazardous gases. For detailed instructions on installing and using the wind guard, refer to the G7 EXO Technical User Manual. | |