Gas Sensor Cross Sensitivity
This article describes gas cross-sensitivity, how it can impact your gas sensors, and which gas sensors can exhibit it. The article applies to G7c, G7x, G8, and EXO.
What is gas sensor cross sensitivity?
Cross sensitivity is a phenomenon that occurs when a gas other than the gas being monitored causes the gas sensor to show a reading, even when the target gas is not present. An example of this would be a CO sensor responding to H₂ in the area.
What does Blackline do to alleviate cross sensitivity?
To alleviate cross sensitivity, Blackline:
- Chooses sensors that are less likely to show a cross sensitive response
- Includes filtration to remove the cross sensitive gases
- Tests sensor and gas combinations to confirm the cross sensitive responses to other gases
What will I see during Calibration or Bump Testing when sensors are cross sensitive?
During calibration and bump testing a cross sensitive sensor, you will see readings on the sensor you are not calibrating or bump testing because the sensor has a response to the gas being applied during the calibration or bump testing.
How can I reduce cross-sensitive response in gas sensors?
To reduce cross-sensitive responses in gas sensors, Blackline strongly recommends following the prescribed order of the gases applied to the device when doing either Dock or manual calibration or bump testing. See Gas Sensor Bump Testing and Calibration Order for more information.
What can I do when sensors get cross sensitive readings?
For most sensor/gas combinations, the cross-sensitive response can be removed by exposing the device to fresh air for 1–2 minutes. If the sensors do not return to zero after fresh-air exposure, contact Technical Support.
What gas combinations produce noticeable cross sensitivity?
Almost all gases cause some cross sensitivity, but our devices are designed to filter out many of the most commonly occurring ones.
The following sensors can have cross sensitivity.
| Sensor | Sensor Code | Notes on Cross-Sensitivities |
|---|---|---|
| Cl2 | L |
|
| ClO2 | W |
|
| CO | M |
|
| COSH (CO + H2S) (North America and International) | D | CO Sensor:
H2S Sensor:
|
| COSH (CO + H2S) (Europe) | 5 | CO Sensor:
H2S Sensor:
|
| CO-H | R |
|
| High-range CO | T |
Note: Activated carbon filter cloth removes SO2, NO2, and H2S, and provides short term (<1000 ppm hours) protection against methanol, ethanol, IPA. |
| High-range H2S | J |
|
| H2 | G |
Note: Continuous high level exposure may reduce the efficiency of the filter material.
|
| H2 (North America) | 7 |
|
| H2S | H |
|
| HCN | N |
|
| LEL-MPS | F |
|
| NH3 | A |
|
| High-range NH3 | B |
|
| NO2 | K |
|
| O2 | O |
|
| O3 | E |
|
| PID | V |
Note: Sensor accuracy decreases with increased relative humidity and temperature, and exposure of the sensor to very humid, acidic (sour) and salty environments. This may cause inorganic salts to accumulate on PID enclosure walls, which ultimately compromises the screening potential of the MiniPID 2 fence electrode. |
| SO2 | S |
|
What can I do if I suspect a gas sensor is poisoned?
Exposure to certain chemicals or environmental conditions can poison a gas sensor. Poisoning permanently affects the sensor's readings. If you suspect a gas sensor is poisoned, calibrate the device to find out.
To check a sensor you suspect is poisoned, do the following:
- Move the Blackline Safety device to a clean-air area.
- Use a clean, dry cloth to wipe the body of the device and cartridge.
- Calibrate the device. For G7 and G8, see Performing Calibrations That Are Not Overdue. For EXO, see Manual Calibration.