Help Center

Search the Help Center

Intrinsically Safe Certification

Intrinsically Safe

This device is certified Intrinsically Safe for use in Class I Division 1 Groups A,B,C,D T3; Ex ia IIC T3 Ga; Class I Zone 0 AEx ia Group IIC T3 Ga hazardous (classified) locations.

MC 267256
UL 60079
Class I Division 1 Groups A,B,C,D; T3
Class I Zone 0 AEx ia IIC T3 Ga
CAN/CSA C22.2 No. 60079
Ex ia IIC T3 Ga

IECEx/ATEX: IECEx SIR 20.0022X; Sira 20ATEX2004X; CSAE

IEC 60079; EN 60079

Ex ia IIC T3 Ga

II 1 G2813

-20°C ≤ Tamb ≤ +50°C

Base unit P/N: EX8b-cc-dde

Where b is the model code and cc-dde is the regional code.

Gas cartridge: Standard P/N "Z" | Single-gas P/N "S-#" | Multi-gas P/N "Q-####" |

Caution

For safety reasons this equipment must be operated and serviced by qualified personnel only. High off-scale readings may indicate explosive concentration.

The equipment shall only be charged when in the non-hazardous area using a charger specifically supplied for use with the unit (for example part number JAC2504L-XX, manufactured by Schauer Battery Chargers), approved as SELV or Class 2 equipment against IEC 60950, IEC 61010-1 or an equivalent IEC standard. The maximum voltage and current from the charger shall not exceed 5Vdc and 25A respectively.

The enclosure is manufactured from Aluminium, magnesium, titanium or zirconium which may be used at the accessible surface of the equipment. In rare cases, ignition sources due to impact and friction sparks could occur. This shall be considered when EXO is being installed in Zone 0 locations for group II level of protection Ga.

Consult with your organization’s safety professional for further information regarding the topic of intrinsic safety and any policies, procedures, facilities, or locations within facilities that may be related to intrinsic safety.

Sécurité intrinsèque

Cet appareil est certifié à sécurité intrinsèque pour l’usage en classe I division 1 groupe A,B,C,D T3; Ex ia IIC T3 Ga; classe I zone 0 AEx ia groupe IIC T3 Ga dans les lieux classés comme dangereux.

Standards:

CAN/CSA C22.2 No. 60079-0: 2019

CAN/CSA C22.2 No. 60079-11: 2014

CSA/C22.2 No. 60079-29-1:17 AMD 1

C22.2 No. 152 - M1984 (R2011)

UL 913, 8th Edition

UL 60079-0: 2019, 7th Edition

UL 60079-11: 6th Edition

ANSI/ISA 12.13.01: 2000

UL 61010-1, 3rd Edition (2012), AMD1:2018

ANSI/UL Std, No. 60079-26, 3rd Edition

EN 60079-0: 2012/A11: 2018

EN 60079-11: 2012

EN 60079-26: 2015

IEC 60079-0.2017, 7th Edition

IEC 60079-11: 2011 6th Edition

IEC 60079-26: 2014-10, 3rd Edition

Attention

Pour des raisons de sécurité, cet équipment doit être utilisé, entretenu et réparé uniquement par un personnel qualifié. Des lectures supérieures à l’échellepeuvent indiquer des concentration explosives.

L’équipement ne doit être chargé que dans la zone non dangereuse à l’aide d’un chargeur spécifiquement fourni pour l’utilisation avec l’appareil (par exemple, la référence JAC2504L-NA, fabriquée par Schauer Battery Chargers) SELV ou Classe 2 selon IEC 60950, IEC 61010-1 ou une norme IEC équivalente. La tension et le courant maximum du chargeur ne doivent pas dépasser respectivement 5Vdc et 25A.

S’il vous plaît consulter professionnel de la sécurité de votre organisation pour de plus amples informations concernant le sujet de la sécurité intrinsèque et les politiques, les procédures, les installations, ou emplacements au sein des établissements qui peuvent être liés à la sécurité intrinsèque.