Using the Wind Guard

When bump testing or calibrating EXO Diffusion in windy conditions, you must use the wind guard to help regulate the flow and concentration of gas delivered to the sensors.

  1. Position the Wind Guard in front of EXO with the narrow ends on top.
  2. Slide the Wind Guard into the front air diffusion vents (underneath the top lights). When the Wind Guard is in the proper position, the two clamps lock.
  3. Bump test or calibrate as usual.
  4. When bump test or calibration is complete, remove the Wind Guard and store it in a clean, dry environment for later use.
    Warning:

    You MUST remove the Wind Guard when the bump test or calibration is complete. Failure to remove the Wind Guard slows down EXO’s response to hazardous gases.