Help Center

Search the Help Center

Performing a Pump Block Test

  1. Open the Main menu.
  2. Select Gas options.
  3. Select Pump block test.
  4. Use the left, right, and center button to select All inlets to perform a pump block test for all inlets.

    Or to perform a block test for a specific inlet, select the inlet.

    EXO begins the pump block test.

    A message displays to indicate when the pump block test is completed.

  5. If a "Block test failed" message displays on EXO's screen, check the tubing for obstructions and try the pump block test again.