Manual Zero Sensors

You can zero a sensor by using the atmosphere to reset the sensor’s baseline reading. You can also apply inert purge gas to the manual calibration inlet to zero sensors.

Warning:

You must zero EXO’s sensors with clean air. If a sensor is zeroed where its targeted gas levels are abnormal, the gas levels EXO displays will not be accurate. Inaccurate readings are a safety hazard.

  1. Ensure EXO is in clean air.
  2. Open the Main menu.
  3. Select Gas options.
  4. Select Zero sensors.
  5. When prompted, ensure the gas sensors you wish to zero are selected.
  6. Select Start zeroing.
    Note:

    Do NOT apply gas.

    A message displays to indicate when the zeroing is completed.

If you see a “Zero incomplete” message on EXO’s screen:

  • EXO may be in an environment with abnormal gas levels.

  • EXO’s cartridge may need to be replaced.