MPS Zero Confirmation
When powered on, devices equipped with an MPS sensor require confirmation that the environment is suitable for automatic zeroing.
- A clean-air confirmation prompt displays. The user must confirm within 15 seconds.
-
G7/G8: Clean-air confirmation prompt displays after the initial start-up sequence.
-
EXO: Prompt displays during the setup wizard unless manually bypassed.
-
- If confirmed, automatic zeroing proceeds.
- If not confirmed, a latched notification triggers. The device must be power cycled before start-up can continue.
If the clean-air confirmation prompt is disabled in the configuration profile:
- The device prompts the user to manually zero the sensor through the Zero Sensors menu.
- Sampling does not start until manual zeroing completes.
MPS sensors must be started in clean air to ensure accurate readings. Blackline recommends an environment with at least 18% oxygen when zeroing.