When powered on, devices equipped with an MPS sensor require confirmation that the environment is suitable for automatic zeroing.

  • A clean-air confirmation prompt displays. The user must confirm within 15 seconds.
    • G7/G8: Clean-air confirmation prompt displays after the initial start-up sequence.

    • EXO: Prompt displays during the setup wizard unless manually bypassed.

  • If confirmed, automatic zeroing proceeds.
  • If not confirmed, a latched notification triggers. The device must be power cycled before start-up can continue.

If the clean-air confirmation prompt is disabled in the configuration profile:

  • The device prompts the user to manually zero the sensor through the Zero Sensors menu.
  • Sampling does not start until manual zeroing completes.
Note:

MPS sensors must be started in clean air to ensure accurate readings. Blackline recommends an environment with at least 18% oxygen when zeroing.