Editing Configuration Profile Details
Configuration profile description:Manage the configuration profile name, device language, and assigned team member display settings.
Operating modes (G7): Configure the functional and sensor settings according to operating mode (normal operation, pre entry, high risk, leak check, pump run, and SCBA). Modes cause devices to exhibit modified behavior depending on the environment or situation.
For detailed information on modes and settings, see Editing Configuration Profile Operating Mode Settings
Functional settings: Manage general device settings. Depending on the device type and operating mode, functional settings allow you to configure the behavior of certain device features (e.g., SOS Alert, device power down, low-battery threshold, or fall detection threshold).
Sensor settings: Manage settings specifically related to gas and gamma detection. Depending on the device type, sensor settings let you enable or disable certain gas sensors, indicate alarm thresholds per sensor type, and specify bump testing and calibration intervals.
The gamma sensor is factory calibrated. Calibration interval settings are not available.
Pump module (EXO): Configure EXO’s pump module. For each inlet, set the Inlet function as gas sampling, self-bump test and calibration, or disabled.
The pump module settings do not apply to the EXO Diffusion module.
You must configure inlets in order, with disabled inlets listed after self-bump test and calibration inlets, purge inlets, and gas sampling inlets.
For inlets configured for gas sampling, toggle the Default to ON upon startup setting to ON or OFF for each inlet. If multiple inlets are used for gas sampling, enter the Sampling schedule (minutes). The interval entered indicates the amount of time EXO pumps gas over the sensors from each inlet. There is a two-minute buffer period in between each sample where air is pumped over the sensor to purge the gas from the first inlet before replacing it with gas from the next inlet.
A multi-inlet sampling setup implies that EXO does not continuously monitor any single environment.
If enabled, you can configure the EXO Pump module to complete automatic bump tests and calibrations. EXO initiates its own bump tests and calibrations when they are overdue.
You must have a service plan to enable automatic bump tests and calibrations. For more information, contact your Customer Relationship Manager (CRM).
You must configure EXO’s pump inlets 1 and 2 to complete automatic bump tests and calibrations. Confirm that one of the inlets configured to self-bump test and calibration functionality is set to purge and is connected to an Ultra Zero Grade Air gas cylinder to run automatic calibrations. If you do not configure an inlet to purge, EXO only performs automatic bump tests.
Interface Ports (EXO): Configure EXO's interface ports to determine how EXO communicates a signal to an external device (e.g., a horn or light). Use the settings to indicate which event type activates each interface port and define how EXO supplies power to external devices when an event occurs.
The interface ports can be activated by a high-gas event, a low-gas event, a text message, or an AlertLink message. You can also configure EXO to apply or remove power from an external device for the duration of an event.
Assigned Devices: Assign devices to the configuration profile by selecting a device from the list. The assigned devices are automatically highlighted.
Last profile change (G6): Display the last time the profile was changed and the name and organization of the person who last edited the profile.
AlertLink messaging (EXO): Configure AlertLink message notification behavior on EXO devices when a message is received by the device. Use the setting to determine if AlertLink messages persist for a maximum of 90 minutes until they are manually acknowledged on the device, or if the message automatically clears after a pre-set timeout period between 5 and 90 minutes (15 minutes is the default).